Medical Techlabs Inc. / 3 May 2019 / Mark Smith
ISO 13485 Audit Checklist
Inspection score
96.35%
Failed items
0
Created actions
1
Organization
Medical Techlabs Inc.
Conducted on

3rd May, 2019

3:30 PM +08

Prepared by
Mark Smith
Location
Stowe Rd Winchester, CA 92596 United States (33.7226518023363, -117.0773956462627)
Actions
1
ISO 13485 Gap Analysis
1
96.35%
Management Controls
1
94.74%
Quality Manual defines scope of QMS, procedures (or reference to) within QMS, and description of the interaction of processes within QMS? (ISO 13485:2016: 4.1, 4.2.2)
Yes
Criteria and methods are in place to monitor and control processes for effectiveness? (ISO 13485:2016: 4.1.3(a), 4.2.1(d), 8.4)
Yes
Firm has established and conducts Management Reviews, at least annually? (ISO 13485:2016: 5.1(d), 5.6)
Yes
Management reviews examine suitability and effectiveness of quality systems, improvements needed because of customer requirements, and resource needs? (ISO 13485:2016: 4.1.3(c), 5.6.1, 5.6.3, 6.1, 8.4)
Yes
Management review addresses audit results, customer feedback, process performance, Corrective and Preventive Actions (CAPAs), previous management reviews, changes to QMS, recommendations for improvement, and new or revised regulatory requirements? (ISO 13485:2016: 5.6.2)
Yes
Notes
I think we can improve on this moving forward not that we have utilized our digital audit tool. We previously depended on paper checklists for audits and we just found we missed audits in the past.
Firm has established a Quality Manual and Quality System Procedures and Instructions that are appropriate? (ISO 13485:2016: 4.1.2(a), 4.2.1(b), (c)
Yes
Firm has established Quality Plan? (ISO 13485:2016: 4.2.1(d), 5.4)
Yes
Quality Planning addresses QMS needs and Quality Objectives? (ISO 13485:2016: 5.4.2)
Yes
Firm has implemented Quality Policy and Quality Objectives? (ISO 13485:2016: 4.2.1(a), 5.1(b), (c), 5.3, 5.4.1)
Yes
Firm has established Quality Audit procedures and conducts audits? (ISO 13485:2016: 4.2, 8.2.4)
Yes
Notes
Will meet with head auditor to check on the new tool we have.
Actions
To do
Update on digital audits
Quality audits examine compliance and effectiveness? (ISO 13485:2016: 4.1.3(c), 4.2.1(d), 8.2.4)
Yes
Auditors are trained? (ISO 13485:2016: 6.2, 8.2.4)
No
Notes
We have a new auditor who is yet to be on boarded and trained. Will discuss this with Angela later.
Audits are conducted by objective parties? (ISO 13485:2016: 8.2.4)
Yes
Quality audits are linked to CAPA? (ISO 13485:2016: 8.2.4)
Yes
Review organizational structure of firm. Resources are available to support processes? (ISO 13485:2016: 4.1.3(b), 5.1(e), 5.5.1, 5.5.2, 6.1, 6.2)
Yes
Firm has defined a management representative with executive responsibility for implementing and reporting quality management system? (ISO 13485:2016: 5.1, 5.5.1, 5.5.2, 6.1, 6.2)
Yes
Appropriate responsibilities, authority, and resources are in place for quality system activities? (ISO 13485:2016: 5.1(e), 5.5.1, 5.5.2, 6.1, 6.2)
Yes
Firm has established procedures for identifying training needs and personnel are trained to perform assigned responsibilities? (ISO 13485:2016: 6.2)
Yes
Executive management ensures adequate and effective quality system is implemented? Ensure management is committed to and communicates importance of meeting customer requirements, regulatory requirements, and QMS. (ISO 13485:2016: 5.1(a), 5.2, 5.5.3)
Yes
Design and Development / Design Controls
100%
Products are subject to design controls? (ISO 13485:2016: 7.1, 7.3)
Yes
Design control and risk management procedures are established and applied? (ISO 13485:2016: 7.3)
Yes
Design and development stages are identified? (ISO 13485:2016: 7.3.2)
Yes
Review, verification, validation, and design transfer activities at each stage are appropriate? (ISO 13485:2016: 7.3.2)
Yes
Responsibilities for design and development are defined? (ISO 13485:2016: 7.3.2)
Yes
Review the project design & development plan, responsibilities, and interfaces. Design & development plan is updated, reviewed, and approved? (ISO 13485:2016: 7.3.2)
Yes
Design input requirements were established, reviewed, and approved? (ISO 13485:2016: 7.2.1, 7.3.3)
Yes
Customer requirements are captured? (ISO 13485:2016: 7.2.1, 7.3.3)
Yes
Inputs include functional, performance, safety, and statutory and regulatory requirements? (ISO 13485:2016: 7.2.1, 7.3.3)
Yes
Incomplete, ambiguous, and/or conflicting requirements were addressed? (ISO 13485:2016: 7.3.3)
Yes
Design & development outputs are established, verifiable, reviewed, and approved? (ISO 13485:2016: 7.3.4(a), (c))
Yes
Design & development outputs are appropriate for purchasing, production, and servicing? (ISO 13485:2016: 7.3.4(b))
Yes
Essential design & development outputs are identified? (ISO 13485:2016: 7.3.4(d))
Yes
Acceptance criteria is referenced by design & development outputs and was defined prior to design verification and design validation activities? (ISO 13485:2016: 7.3.4(c), 7.3.6)
Yes
Design verification confirmed design outputs met design input requirements? (ISO 13485:2016: 7.3.6)
Yes
Design validation results prove device met predetermined user needs and intended uses? (ISO 13485:2016: 7.3.7)
Yes
Design validation did not leave unresolved discrepancies? (ISO 13485:2016: 7.3.7)
Yes
If required by national or regional regulations, confirm clinical evaluations and/or evaluation of device performance were performed? (ISO 13485:2016: 7.3.7)
Yes
If device contains software, confirm software was validated? (ISO 13485:2016: 7.3.2, 7.3.7)
Yes
Initial production units (or equivalents) were used for design validation? (ISO 13485:2016: 7.3.7)
Yes
Risk management activities were performed? (ISO 13485:2016: 7.1)
Yes
Design changes were controlled and validated (or where appropriate, verified)? (ISO 13485:2016: 7.3.2, 7.3.6, 7.3.9)
Yes
Design changes have been reviewed for effect on components and product previously made? (ISO 13485:2016: 7.3.2, 7.3.6, 7.3.9)
Yes
Design reviews were conducted at appropriate stages of design & development? (ISO 13485:2016: 7.2.2, 7.3.2, 7.3.5)
Yes
Design review attendees were appropriate for stage and included independent reviewer? (ISO 13485:2016: 7.3.2, 7.3.5)
Yes
Design was correctly transferred to production? (ISO 13485:2016: 7.3.2, 7.3.8)
Yes
DHF contains design control documentation? (ISO 13485:2016: 7.3.10)
Yes
Corrective and Preventive Actions (CAPA)
87.5%
CAPA procedures comply with regulatory requirements? (ISO 13485:2016: 4.1, 4.2, 8.5)
Yes
Non-conforming product and CAPA procedures determine the need for investigation and notification? (ISO 13485:2016: 8.3, 8.5)
Yes
Non-conforming product and CAPA procedures define responsibilities for review and disposition? (ISO 13485:2016: 8.3, 8.5)
Yes
Procedures for rework, retesting, and re-evaluation of nonconforming product exist and are followed? (ISO 13485:2016: 8.3, 8.5)
Yes
Appropriate records of quality problems have been created and used? (ISO 13485:2016: 8.3, 8.5)
Yes
Trend analysis data indicates quality problems? Data used for CAPA decisions? 9ISO 13485:2016: 8.1, 8.2.5, 8.4, 8.5)
Yes
CAPA data is complete, accurate, and timely? Compare results across multiple data sources to identify quality problems. (ISO 13485:2016: 8.4, 8.5)
No
Notes
Our record keeping used to be filled on paper and it had caused delays in reporting and action. Tagging this as no for now and until I am satisfied with the meeting later will I change to yes.
Appropriate statistical techniques are implemented? (ISO 13485:2016: 8.1, 8.2.5, 8.4)
Yes
Device failure investigations determine root cause? (ISO 13485:2016: 8.3, 8.5)
Yes
Failure investigations are commensurate with risks? (ISO 13485:2016: 8.3, 8.5)
Yes
Controls exist to prevent non-conforming product from being released? (ISO 13485:2016: 8.3)
Yes
Appropriate actions were taken for quality problems? (ISO 13485:2016: 8.2.5, 8.5.2, 8.5.3)
Yes
CAPA actions were effective, verified, validated, documented, and implemented appropriately? (ISO 13485:2016: 8.5)
No
CAPAs and nonconformities were disseminated to personnel responsible for ensuring quality and prevention of problems? (ISO 13485:2016: 8.3, 8.5)
Yes
Quality issues and CAPAs were disseminated for Management Review? (ISO 13485:2016: 5.6.3, 8.3, 8.5)
Yes
Firm has procedures for handling complaints and investigation of advisory notices / recalls? Ensure provisions exist to feed in CAPA system. (ISO 13485:2016: 7.2.3, 8.2.1, 8.2.2, 8.2.3)
Yes
Medical Device Reporting (MDR)
100%
Medical Device Reporting (MDR) procedures comply with regulatory requirements? (ISO 13485:2016: 8.5.1)
Yes
Firm maintains MDR event files that comply with regulatory requirements? (ISO 13485:2016: 8.5.1)
Yes
Appropriate MDR information is identified, reviewed, reported, documented, and filed? (ISO 13485:2016: 8.5.1)
Yes
Firm is effective in identifying MDR reportable events? (ISO 13485:2016: 8.5.1)
Yes
Firm has established procedures for receiving, reviewing, and evaluating complaints? (ISO 13485:2016: 7.2.3, 8.2.1, 8.2.2, 8.5.1)
Yes
Firm maintains complaint files and that they are reasonably accessible? (ISO 13485:2016: 7.2.3, 8.2.1, 8.2.2, 8.5.1)
Yes
Complaints are evaluated to determine if an event should be a MDR? (ISO 13485:2016: 7.2.3, 8.2.1, 8.2.2, 8.2.3, 8.5.1)
Yes
Complaint investigations include the device name, date of complaint, device identification number, contact information of complainant, details of complaint, date and results of investigation, any corrective actions, and replies to complainant? (ISO 13485:2016: 7.2.3, 8.2.1, 8.2.2, 8.5.1)
Yes
Production and Process Controls (P&PC)
100%
Product realization processes are planned? Confirm that risk management occurs throughout product realization. (ISO 13485:2016: 7.1)
Yes
Planning of product realization is consistent with requirements of other processes of QMS? (ISO 13485:2016: 7.1)
Yes
Requirements have been defined for suppliers, contractors, and consultants; ensure suppliers, contractors, and consultants are selected on ability to meet requirements? (ISO 13485:2016: 7.1, 7.4.2)
Yes
Firm maintains records of acceptable suppliers, contractors, and consultants? (ISO 13485:2016: 7.4.1)
Yes
Data supporting supplier requirements is maintained? Verify that suppliers, contractors, and consultants agree to notify firm of changes in products and/or services. (ISO 13485:2016: 7.4)
Yes
Procedures for identifying product during all stages of receipt, production, distribution, and installation are in place? (ISO 13485:2016: 7.5.8, 7.5.9)
Yes
Firm maintains procedures and records for traceability of each unit, lot, or batch of finished devices and components? NOTE: may not be required for all devices. (ISO 13485:2016: 7.5.9)
Yes
Process is controlled and monitored? (ISO 13485:2016: 7.5, 7.6, 8.2.5, 8.2.6, 8.4)
Yes
Equipment used has been adjusted, calibrated, and maintained? (ISO 13485:2016: 7.5)
Yes
Control and oversight activities identified? Ensure control of inspection, measuring , test equipment, and calibration. (ISO 13485:2016: 7.6, 8.4)
Yes
Firm has established procedures for production and process changes? Ensure changes are verified or validated, as needed. (ISO 13485:2016: 7.3.9, 7.5.6)
Yes
Device history record (DHR) to identify rejects and/or non-conformances? (ISO 13485:2016: 8.3)
Yes
Defects, rejects, non-conformances, and removal of materials were handled properly? (ISO 13485:2016: 8.3)
Yes
Processes that cannot be fully verified are validated? (ISO 13485:2016: 7.5.6)
Yes
Automated or software driven processes are validated for intended uses? (ISO 13485:2016: 7.5.6)
Yes
Validations are documented and conducted by qualified personnel? (ISO 13485:2016: 7.5.6)
Yes
Personnel are trained per manufacturing processes and aware of potential defects? (ISO 13485:2016: 6.2)
Yes
Monitoring and control methods, data, date performed, individuals performing the process, and the major equipment used is documented? (ISO 13485:2016: 7.1, 8.4)
Yes
Linkages to other processes determined? (ISO 13485:2016: 4.1, 4.2)
Yes
Infrastructure and work environment are appropriate and controlled? (ISO 13485:2016: 6.3, 6.4)
Yes
Maintenance schedules, routine inspections, and adjustments to equipment occur? (ISO 13485:2016: 6.3, 7.5.1, 7.5.6, 7.6)
Yes
Procedures are in place for contamination control and cleanliness? (ISO 13485:2016: 6.4.2, 7.5.2)
Yes
Verification of purchased products is adequate? (ISO 13485:2016: 7.4.3, 8.4)
Yes
Procedures define receiving, in-process, and final acceptance activities? (ISO 13485:2016: 7.5.11, 8.4)
Yes
Receiving, in-process, and final acceptance activity records exist? (ISO 13485:2016: 8.4)
Yes
Procedures exist and that acceptance status of product is indicated? (ISO 13485:2016: 7.1, 8.2.6)
Yes
Procedures define labeling activities, including integrity, inspection, storage, operations, and control numbers? (ISO 13485:2016: 7.5.11)
Yes
Product packaging and shipping containers adequately protect device during processing, storage, handling, shipping, and distribution? (ISO 13485:2016: 7.5.11)
Yes
Notes
There was a shipping issue discovered by our auditor last week. I saw the photos on the latest report this morning and tagging this as yes.
Procedures exist to prevent mix-ups, damage, deterioration, contamination, or other adverse effects to product during handling? (ISO 13485:2016: 7.5.11)
Yes
Notes
We have incorporated the photo option in our quality check for shipping and it helped me greatly.
Procedures exist for product distribution; confirm distribution records include name and address of consignee, identification and quantity shipped, date of shipment, and identification numbers? (ISO 13485:2016: 4.2.3, 7.1, 7.5.8, 7.5.9.2, 7.5.11)
Yes
Installation and inspection procedures exist (if applicable)? Verify installation records are maintained. (ISO 13485:2016: 7.5.3)
Yes
Servicing procedures exist (if applicable)? Verify servicing records are maintained. (ISO 13485:2016: 7.5.4)
Yes
Firm identifies, verifies, protects, and safeguards customer property under its care? (ISO 13485:2016: 7.5.10)
Yes
Sterilization Process Controls
100%
Review sterilization process procedures. Sterilization process is validated? (ISO 13485:2016: 7.5.7)
Yes
Review sterilization control and monitoring activities. Processes, equipment, and calibration are current? (ISO 13485:2016: 7.5.5)
Yes
Purchasing Controls
60%
Suppliers are evaluated for ability to meet specified requirements? (ISO 13485:2016: 7.4.1)
Yes
Adequacy of specifications of materials and/or services provided by supplier is confirmed? (ISO 13485:2016: 7.4.2)
Yes
Purchasing information identifies requirements for approval of product, procedures, processes, and equipment, requirements for personnel qualification, and QMS requirements? (ISO 13485:2016: 7.4.2)
Yes
Supplier evaluation records are maintained? (ISO 13485:2016: 7.4.1)
No
Notes
Our old supplier evaluation records are missing in the filing cabinet but I asked Mary last week to get copies. I am hoping she has them later. What I can assure is moving forward it will not be an issue because of our new tool.
Photos
Verification and acceptance of purchased materials and/or services is adequate? (ISO 13485:2016: 7.4.3)
No
Documentation and Records
100%
Documents and changes are approved prior to use? (ISO 13485:2016: 4.2.4)
Yes
Documents and records are legible and identifiable? (ISO 13485:2016: 4.2.4(e), 4.2.5)
Yes
Documents of external origin are identified with controlled distribution? (ISO 13485:2016: 4.2.4(f))
Yes
Firm maintains a quality system record (QSR) which includes or refers to location of procedures? (ISO 13485:2016: 4.2.1(c), (e))
Yes
Documents and records are retained for required length of time (this includes retention of obsolete controlled documents and records)? (ISO 13485:2016: 4.2.1, 4.2.4, 4.2.5)
Yes
Change records are reviewed and approved by the same functions that performed original review and approval? (ISO 13485:2016: 4.2.4, 7.3.9)
Yes
Change records include a description of change, identification of affected documents, approval signatures, approval date, and effective date? (ISO 13485:2016: 7.3.9)
Yes
Documents are available at point of use and obsolete document are not in use? (ISO 13485:2016: 4.2.4(d), (h))
Yes
Firm maintains DMRs for each type of device? (ISO 13485:2016: 4.2.1)
Yes
DMRs contain or make reference to device specifications, production process specifications, quality assurance procedures and specifications (including acceptance criteria), packaging and labeling specifications (including acceptance criteria), and installation, maintenance, and servicing procedures? (ISO 13485:2016: 4.2.1)
Yes
DHRs are maintained and devices are manufactured according to DMR? Realization processes and product meet requirements? (ISO 13485:2016: 7.1, 8.2.6)
Yes
DHRs contain or make reference to dates of manufacture, quantity manufactured, quantity released for distribution, acceptance records demonstrating the device was manufactured per DMR, primary identification label and labeling used for each unit, and device identification and/or control numbers used? (ISO 13485:2016: 8.2.6)
Yes
Firm maintains records for education, training, skills, and experience of resources? (ISO 13485:2016: 6.2 (e))
Yes
Firm maintains purchasing and supplier records? (ISO 13485:2016: 7.4.1, 7.4.3)
Yes
Sterilization process parameters and records are maintained for each batch? Ensure sterilization validation records are maintained. (ISO 13485:2016: 7.5.5, 7.5.7)
Yes
Customer Requirements
100%
Review product requirements. The intended use, customer requirements, and regulatory requirements are addressed? (ISO 13485:2016: 7.2.2)
Yes
Incoming contracts and orders are reviewed to resolve conflicting information and that customer requirements can be met? (ISO 13485:2016: 7.2.2)
Yes
Procedures and systems exist for customer communications and feedback? Ensure integration with CAPA system. (ISO 13485:2016: 7.2.3, 8.2.1)
Yes
Technical Files
100%
Review technical file procedures (ISO 13485:2016: 4.2.1(d))
Yes
Documents need to ensure planning, operation, and control of technical file processes? (ISO 13485:2016: 4.2.1(d))
Yes
Documentation addresses a general description of product, intended use(s), and any variants, accessories, or other devices used in combination with product? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Design specifications, standards applied, and results of risk analysis are present? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Principal requirements have been fulfilled? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Techniques used to verify design and validate product(s) clinical data? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Documentation defines sterilization method and validation? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Documentation includes instruction manual(s) and labeling? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Major subcontractors have been documented? (ISO 13485:2016: 7.1, 7.2, 7.3.4)
Yes
Completion
Comments/Recommendations
We have an opportunity to improve our record keeping after the incident of the missing cabinet with our files last quarter. Apart from that, we are almost ready for the third party certification. I will meet with Mary today to discuss how’s the auditing experience with the new iAuditor tool. All audit and inspection results are now in the cloud and I am confident we will not have any records missing again with this new tool.
Name and signature
Mark Smith
3rd May, 2019 4:02 PM +08